data from 31.05.2022

ROI0_Re_Measure22.dat
Cs Raman pi pulse with RB2 fidelity
Square pulse using end of DDS playback, 15.784us pulse duration
DExTer has 30us time step length to account for off time within DDS playback
RB1 is on in between the pi pulses for 2us.

1. RSC
2. sequence of pi pulses
3. pushout


ROI0_Re_Measure24.dat
Cs pushout fidelity
Tested and a 1ms ramp down (shutter open step) is sufficient to pushout population in F=4
So using 1ms ramp down for pushout 500us pulse and then 1ms ramp back up
Ramping down to 0.23V (this is after RSC)

1. RSC
2. Raman pi pulse transfers to F=3
3. sequence of ramp trap down 1ms, pushout pulse 500us, ramp up 1ms


data from 01.06.2022
ROI1_Re_Measure1.dat
Rb pushout fidelity

81 us MW pi pulse/ Then 3ms for opening probe shutter where trap ramps to 2.05 V.
Then pushout sequence that is repeated is :
1ms power ramp down to 0.5V
0.5ms push
1ms ramp back up to 2.05V

1. RSC
2. MW pi pulse transfers to F=1
3. pushout sequence

data from 02.06.2022
ROI1_Re_Measure8.dat
Rb MW pi pulse fidelity. Use 85us for pi pulse.
MW frequency 6844.7700(4) MHz
Ramp trap down to 0.2V = 0.3mW

1. RSC
2. series of MW pi pulses 
3. pushout